Series 4000

Series 4000

Multi-Component Gas Mixing System

A low cost, multi-component gas mixing system with an advanced Windows user interface, the Series 4000 now has the expanded capability of mixing up to seven individual gases in a balance gas. The standard unit also comes with USB.

Download Full Datasheet for Series 4000 (pdf)

The Environics® Series 4000 is a multi-component gas mixing system that now automatically mixes up to seven individual gases in a balance gas. 

The gas mixes can be used in generating precise gas calibration standards, creating gaseous atmospheres or producing gas mixes for analytical research or production purposes. The Series 4000 can produce gas concentrations from percent to ppb levels for single or multi-point calibration.

The system consists of two components: the Series 4000 instrument and the user's personal computer. The user interface is a Microsoft® Windows application that communicates with the Environics system via a USB (standard) or RS232 serial interface. Bluetooth enabled devices running Windows software can also control the unit remotely (optional).

Additional options include the ability to remote start, remote stop, and canister filling stop all via a dry contact closure. Optional alarms available and can include visual alarms upon low input cylinder pressure.

//www.environics.com/wp-content/uploads/2020/07/4000_front_view-resized-600.png

Units can now be ordered or retrofit to run on a 24V power source so you can use your unit anywhere.

//www.environics.com/wp-content/uploads/2020/07/Screenshot-2020-07-09-at-11.37.02.png

The Series 4000 consists of a single chassis now supporting up to eight mass flow controllers. The mass flow controllers are factory calibrated using a primary flow standard traceable to the United States National Institute of Standards and Technology (NIST). Each flow controller utilizes an 11 point calibration table with linear interpolation, to increase accuracy and reduce flow controller nonlinearity.

All units are built to your specific needs and can include additional features such as cleaning for oxygen service, Silconert/Sulfinert coating, speciality seals (Kalrex, metal). Custom systems can also be built to handle explosive gases.

  • Simultaneously mixes up to SEVEN gases in a balance gas at user set concentrations.
  • Allows multi-point calibration of analyzers.
  • Automatic calculation of dilution and span gas flows.
  • Internally-stored Mass Flow Controller calibration tables.
  • Modular design allows for additional gas circuits to be added.
  • User definable cylinder library allows for easy selection of frequently used gas cylinders. Each cylinder may contain an unlimited number of component gases, with automatic K-factor calculation.
  • Easy to operate software included. How-to videos available.
  • DLL Drivers Available
The Series 4000 has three basic modes of operation.
  • Concentration Mode Allows user to create a blend by entering target gas concentrations for each source gas cylinder, and the desired total output flow for the mix.
  • Flow Mode Allows user to specify the flow rate of each gas cylinder.
  • Program/Schedule Modes Provides the ability to program the instrument for unattended operation. Programs can be recalled and run in any sequence, at various times/dates.
  • Purge Mode (optional) Allows user to flush the gas flow path from a designated input port. User can define the flow rates used and can purge any combination of component MFCs.
Performance*
  • Accuracy
  • Concentration: ± 1.0% setpoint
  • Flow: ± 1.0% setpoint
  • Flow Repeatability: 0.05%

  • Warm up time: 30 minutes

  • Performance specifications are valid when all Mass Flow Controllers are operating between 10% and 100% of full scale flow. Mass flow controllers are calibrated using a NIST traceable Primary Flow Standard, using a Reference Temperature of 0°C (32°F) and a Reference Pressure of 760 mm Hg (29.92 in. Hg).

Mechanical
  • Gas connections: 1/4" Swagelok (or compatible fitting)

  • Operating Pressures (flow dependent)
  • Minimum: 10 psig (0.67 Bar)
  • Recommended: 25 psig (1.68 Bar)
  • Maximum: 75 psig (5.04 Bar)

  • Wetted Surfaces
  • Tubing: Electropolished 316 Stainless Steel
  • MFCs: Stainless Steel
  • Seals: Viton®
  • (Optional: Kalrez®, Buna-N, Neoprene, Metal)

  • Operating temperatures -32° - 104° F (0° - 40° C)

  • Performance temperatures - 15° - 35° F

  • Weight
  • 2 MFC System: 18 lbs.
  • 4 MFC System: 23 lbs.

  • Dimensions (w x h x d)
  • Portable: 17" x 7" x 15"
  • (43.18 cm x 17.78 cm x 38.1 cm)

  • Rack: 19" x 7" x 15"
  • (48.26 cm x 17.78 cm x 38.1 cm)

Electrical
  • 100 to 250 VAC, 50/60 Hz; 24V now available

Electronics
  • 12 Bit A/D and D/A Conversion
  • USB (standard) or RS232 Serial interface

Software
  • Environics Instrument Control Software (supplied on thumbdrive or via download). How-to videos available.
  • DLL Drivers Available

PC Requirements
  • IBM PC or compatible (486-33 or higher)
  • Windows 7/8/10
  • 1GB RAM
  • 100 MB Hard Disk Space
  • USB (standard) or RS-232 Communication port

  • Viton® and Kalrez® are registered trademarks of Dupont Dow Elastomers
  • Additional input ports (up to 8)
  • Additional MFC's (up to 8)
  • Various Seal Materials including Teflon and Kalrez
  • Cleaned for Oxygen Service
  • Humidification
  • Permeation
  • Silicon Coated Gas Path
  • Purge Circuit
  • Heated Flow Path
  • Canister Filling
  • Explosive Gas system